LAMBDA System 2: -55°C to +220°C

LAMBDA System 2: Thermal conducitvity of fluids from -55°C to +220°C.
LAMBDA System 2: Thermal conducitvity of fluids from -55°C to +220°C.

The LAMBDA System measures the thermal conductivity of fluids according to ASTM D 2717. It is also very suitable to measure the thermal conductivity of fluids and dispersions containing nanoparticles (nanofluids).

The sample temperature is controlled by a special double-jacket heating/cooling device which provides a homogeneous temperature distribution. It can be connected to any circulation thermostat.

Further Characteristics:

  • Thermostats for temperatures from -55°C to +220°C available
  • Temperature ramps and sweeps possible
  • Measurements under pressure and inert gas flushing possible
  • Windows software for control and data visualisation

Applications:

  • Optimization of fluid application and development
  • Quality control
  • Nanofluids

[Data Sheet...]

[Contact...]


Technical Properties:

  • Measuring of: Fluids (also with nano particles), powders.
  • Measuring range: 10 mW/mK to 2000 mW/mK  (other ranges available)
  • Reproducibility: 1%
  • Measuring frequency: every 60 sec approx.
  • Fluid viscosity: 0.1 to 100,000 mPas
  • Pressure range: ambient or 35 bar (500 psi)
  • Temperature range: -55 to +220°C (depends on thermostat), resolution: 0.1°C, accuracy: 0.1°C
  • Temperature measuring: Pt100 (4-wire)
  • Heating power: max. 650 W
  • Power consumption: Instrument max. 50 W, 240/120V at 50/60Hz (wide range input)
  • Sensor power consumption: < 10mW
  • Data transfer: RS 232, analisys program requires Win95/98/NT/XP/2000